IEC/TS 62736 Ed. 2.0:2023 (en)
超音波-パルスエコースキャナ-画像システムの基本性能の安定性を検証するための定期試験の簡易方法
Ultrasonics - Pulse-echo scanners - Simple methods for periodic testing to verify stability of an imaging system's elementary performance

発行年月日: 2023-01-13
状態: 有効
邦訳版: 無

規格概要 IEC TS 62736:2023 is available as IEC TS 62736:2023 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.

IEC TS 62736:2023 specifies requirements and methods for periodic testing of the quality of diagnostic medical ultrasound systems using reflection-mode (pulse-echo) imaging. Image measurement and interpretation workstations are included. This document includes minimum sets of such tests intended for frequent users of medical ultrasound systems, for quality assurance professionals in their organizations, or those hired from other quality-control and/or service-provider organizations. The procedures are for a wide range of more common diagnostic ultrasound systems, currently operating from 1 MHz to 40 MHz, although available phantoms meet the specifications only from 1 MHz to 23 MHz.
The tests are defined in three levels, with the simplest and most cost-effective performed most frequently:
- Level 1 comprises five quick tests/observations to be performed daily to monthly by those normally operating the systems.
- Level 2 includes one necessary test for all systems in addition to those of Level 1, two Level 1 tests performed more rigorously, two tests that are for special situations or equipment, and one that is just optional, included because it is highly developed. Level 2 tests are performed annually by those with meaningful quality assurance training.
- Level 3 extends the two special situations tests to all systems, adds one optional test and includes a periodic review of the QA programme.
Frequent distance-measurement accuracy tests are recommended in this document only for certain classes of position encoding that are not now known to be highly stable and without bias. QA in all dimensions is recommended in this document as the first test for such systems.